Replaceable graphite electrode An easy-to-replace graphite electrode with a sturdy, snap-in connector means there are no pins to bend or break. Pocket clip A pocket clip is featured on the back. Specifications HI983124 EC Range 0.00 to 20.00 mS/cm Resolution 0.01 mS/cm Accuracy ±2% F.S. TDS Range 0.00 to 10.00 ppt (g/L) Resolution 0.01 ppt (g/L) Accuracy ±2% F.S. Temperature Range 0.0 to 60.0°C / 32.0 to 140.0°F Resolution 0.1°C / 0.1°F Accuracy ±0.5°C / ±1°F Additional Specifications Calibration automatic, one point at 12.88 mS/cm or 6.44 ppt (g/L) TDS Conversion Factor adjustable from 0.45 to 1.00 Temperature Compensation automatic, with β adjustable from 0.0 to 2.4% / °C Probe HI73311 replaceable EC/TDS graphite electrode (included) Environment 0 to 50°C (32 to 122°F); RH max 100% Battery Type / Life 1.5V (4) / approx. 100 hours of continuous use; auto-off after 8 minutes of non-use Dimensions 163 x 40 x 26 mm (6.4 x 1.6 x 1.0’’) Weight 100 g (3.5 oz.) Ordering Information HI983124 is supplied with HI73311 EC/TDS probe, HI73128 probe removal tool, 12880 µS/cm conductivity standard sachets (3), 6.44 ppt TDS standard sachets (3), batteries, and instructions. HI983124 EC/TDS and Temperature Tester • Waterproof · Waterproof and designed to float • Automatic Temperature Compensation (ATC) • HOLD feature · HOLD button to freeze readings on the display • Battery Error Prevention System (BEPS) · Alerts the user of low battery power that could adversely affect readings EC and TDS solutions begin on page 5.34; see EC electrode specifications on page 1.70 1 Testers and Monitors 1.42 | www.hannainst.com testers

RkJQdWJsaXNoZXIy MTI3NTM=